Joan Laboratory P20 P200 Micropipette Autoclavable Technical Data
PictureModelVolume range

Increment

(ul)

Test volume

(ul)

Max permissible

systematic error

(Inaccuracy)

Max permissible

random error

(Imprecision)

%ul%ul
8.jpgM8-100.5 – 10ul0.1

10

5

1

1.50

2.50

4.00

0.15

0.125

0.04

1.50

2.50

4.00

0.15

0.125

0.04

M8-505-50ul0.5

50

25

5

0.50

1.00

2.00

0.50

0.375

0.15

0.50

1.00

2.00

0.25

0.25

0.10

M8-30050-300ul5

300

150

50

0.25

0.50

0.80

2.1

1.5

0.75

0.25

0.50

0.80

0.75

0.75

0.4

12.jpgM12-100.5-10UL0.1

10

5

1

1.50

2.50

4.00

0.15

0.125

0.04

1.50

2.50

4.00

0.15

0.125

0.04

M12-505-50ul0.5

50

25

5

0.50

1.00

2.00

0.5

0.375

0.15

0.50

1.00

2.00

0.25

0.25

0.10

M12-30050-300ul5

300

150

50

0.25

0.50

0.50

2.1

1.5

0.75

0.25

0.50

0.80

0.75

0.75

0.40